The objective of this study was to evaluate a novel accelerometer-based sensor system, the Walkabout Portable Gait Monitor (WPGM), for use in kinetic gait analysis of dogs. The accelerometer was compared to the common reference standard of force platform analysis. Fifteen client-owned, orthopedically sound dogs of various breeds underwent simultaneous force platform and accelerometer gait trials to measure peak vertical forces (PVFs). The agreement between PVF for the accelerometer and force platform was measured using concordance correlation coefficient (CCC) and was found, overall, to be moderate [CCC = 0.51; 95% confidence interval (CI): 0.46 to 0.56]. The agreement between PVF for the accelerometer and force platform for the forelimbs was positive and substantial (CCC = 0.79; 95% CI: 0.74 to 0.84) and for the hind limbs was positive and low (CCC = 0.34; 95% CI: 0.29 to 0.38). As measured by the accelerometer, PVF was systematically higher than as measured by the force platform (forelimbs 55.3 N, hind limbs 144.3 N). It was also found that, when positioned over the lumbar spine, the WPGM cannot measure PVF of the individual forelimbs and hind limbs, which limits its use as a clinical tool to measure kinetic variables in dogs.
Evaluation of a novel accelerometer for kinetic gait analysis in dogs.
Date
2014 Jul
Journal
Can J Vet Res
Volume
78
Number
3
Pages
226-32