Effect of headless compression screw on construct stability for centre of rotation and angulation-based levelling osteotomy. Biomechanical testing in flexion and torsion

Journal
Meyer MR, Perez MA, Hossain MS, Silverman EB, Fitch RB, Wicker RB. Vet Comp Orthop Traumatol. 2017 Jul 20; 30 (4):243-247.

OBJECTIVE: To compare the biomechanical properties of bone and implant constructs when used for the centre of rotation and angulation (CORA) based levelling osteotomy, with and without implantation of a trans-osteotomy headless compression screw tested under three-point flexural and torsional forces; thereby determining the contribution of a trans-osteotomy headless compression screw with regards to stability of the construct.

METHODS: Experimental biomechanical study utilizing 12 pairs of cadaveric canine tibias. Using the CORA based levelling osteotomy (CBLO) procedure, the osteotomy was stabilized with either a standard non-locking CBLO bone plate augmented with a headless compression screw (HCS) or a CBLO bone plate alone. Tibial constructs were mechanically tested in three-point craniocaudal flexural testing or in torsion.

RESULTS: In three-point flexural testing, the difference between the two constructs was not significant. In torsion, the difference in the angle of failure between constructs with a HCS (48.46°) and constructs without a HCS (81.65°) was significant (p = 0.036). Maximum torque achieved by constructs with a HCS (21.7 Nm) was greater than those without (18.7 Nm) (p = 0.056). Stiffness differences between both groups in torsion and bending were not significant. Use of a HCS did increase the stability of the CBLO construct in torsional testing, but not in flexural testing.